APPENDIX E
Register Type D: Units
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w
Bits
HEX
Meaning (binary)
0, 1
0003
00=°C, 01=°F, 10=K
2, 3
000C
Ext press unit, 00=psia, 01=bara, 10=kpa
4
0010
Field cal type, 0=1 gas (offset), 1=2 gas
5
0020
Field cal type 2, 0=1 gas (offset), 1=2 gas
6
0040
Field cal type 3, 0=1 gas (offset), 1=2 gas
7
0080
Date format 0=dd/mm/yy, 1=mm/dd/yy
11, 12, 13
3800
Ext Sensor Parameter (000=none, 001=dewp, 010=temp,
011=press, 100=other)
14, 15
C000
SPARE
Register Type E: Calibration Table Setting
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w
Bits
HEX
Meaning (binary)
0
0001
Phase Control Input, 0–Internal, 1–External (mA Input)
1, 2
0006
Calibration Table, 0=Table 1, 1=Table 2, 2=Table 3
Register F: -2000.0 to +2000.0
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w
Range = 0 to 4000 represents -2000.0 to +2000.0
Conversion: (RegValue – 20000)/10.0
For external sensor values
Dewpoint: -100/+20°C, 148.0/68.0°F, 173.0/293.0 K
Tempr: -50.0/+100.0°C, -58.0/+212.0°F, 223.0/373.0 K
Pressure: 0.0/44.1 psia, 0.0/3.0 barA, 0.0/304.0 kpa
Register Type G: -200.00 to +200.00
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w r/w
Range = 0 to 40000 represents -200.00 to +200.00
Conversion: (RegValue – 20000)/100.00