QMA601 User’s Manual
72
97449 Issue 3, June 2018
APPENDIX A
Mechanical Specifications
Type
GUB Flameproof Exd
Enclosure
Lid & Body
Glass Window
Cast copper-free aluminum LM25 (EN AC-42000), less than 0.6
magnesium
Heat resistant, explosion proof, polyester coated, IP66, NEMA 4
Analyzer Gas Connections
1/8” NPT
Weight
35kg (77lbs) without sampling system
Sample System Enclosure
316L stainless steel
Hazardous Area Specifications
Hazardous Area
Certification
ATEX
IECEx
CSAc
CSAus
TC-TR
II 2 GD Ex d IIB+H2 T6 Gb, Ex tb IIIC T85°C Db, Tamb -40°C
to +60°C
Ex d IIB+H2 T6 Gb, Ex tb IIIC T85°C Db, Tamb -40°C to
+60°C
Class I, ZONES, Ex d IIB+H2 T6 Gb, Tamb -20°C to +55°C
Class I, Div 1, Group BCD T6, Tamb -25°C to +55°C
1Ex d IIB+H2 T4 Gb X, 1Ex tb IIIC 130°C Db X
1Ex d IIB+H2 T3 Gb X, 1Ex tb IIIC 195°C Db X
*Applies only to standard 30-30 second cycle variant. Contact Michell
Instruments for further details.