MAINTENANCE
If the contamination trap is to be returned to Michell Instruments, or a Michell
Instruments' approved service center form, the F0121 Decontamination Certificate must
be completed and sent with the returned trap.
4.4
Removal and Replacement of the Desiccant Column
Removal
1.
Isolate the sample gas supply and de-pressurize the QMA601 sampling
system.
NOTE: Goggles should be worn at all times when working
with pressure.
2.
When the sample system has been depressurized, use the spanner/wrench
sizes indicated below to disconnect the VCR fittings on the Desiccant
Column.
15/16" wrench
3/4" wrench
1/1-16" wrench
5/8" wrench
3.
Pull and twist the index plungers to the unlock position.
Index Plungers