QMA601 User’s Manual
Michell Instruments
71
APPENDIX A
Appendix A
Technical Specifi cation
Performance
Measurement technology
Fast Response Quartz Crystal Microbalance
Calibrated range*
0.1 to 700 ppm
V
Measurement range
0.1 to 2000 ppm
V
Accuracy
±10% of reading from 1 to 2000 ppm
V
±0.1 ppm
V
between 0.1 & 1 ppm
V
Repeatability
±5% of the reading from 1 to 2000 ppm
V
±0.1 ppm
V
between 0.1 & 1 ppm
V
Limit of Detection
0.1 ppm
V
Available Units
ppm
V
, ppm
W
, mg/Nm
3
, vapor pressure (Pa), dew point (°C/°F), lbs/
MMscf
Response Speed
T63 <2 mins to step change in either direction
T95 <5 mins to step change in either direction
Automatic calibration
Internal moisture generator source calibrated traceable to NPL & NIST
Sensitivity
0.01
ppm
V
or 1% of reading, whichever is greater
Electrical Specifications
Supply Voltage
85 to 264 V AC, 47/63Hz or 24 V DC (ATEX / IECEx)
24 V DC only (cCSAus)
Alarms
1 x System Alarm, volt-free change-over (FORM C)
3 x process alarms, selectable for various parameters, volt free
change-over (FORM C)
Analog signals
2 X 4-20mA or 1-5 V (selectable)
Maximum load resistance 500
Ω
for 4-20mA and minimum load of 1M
Ω
for 1-5 V
Digital communications
RS485 Modbus RTU
Modbus TCP
Data logging
Available on analyzer (Limited number of values) or via Application
Software
Local Interface
7” color LCD with intuitive HMI
Electrical Connections
M20 entries for cable glands
Operating Conditions
Inlet Pressure
3 barg (43.5 psig)
Outlet Pressure
2 barg (29 psig)
Sample Flow
300ml/min total flow
Sample Gas Temperature
0 to +100°C (+32 to + 232°F)
Operating Environment
Analyzer only
Analyzer in sampling
System
+5 to +45°C (+41 to +113°F) up to 90%RH
-20 to +55°C (-4 to +131°F) up to 95%RH, (fitted with heater/
thermostat and/or enclosure cooling as appropriate to ma5 to
+45°C internal temperature)
Summary of Contents for QMA601
Page 1: ...QMA601 Process Moisture Analyzer User s Manual 97449 Issue 3 June 2018...
Page 80: ...QMA601 User s Manual 70 97449 Issue 3 June 2018 APPENDIX A Appendix A Technical Specifications...
Page 83: ...QMA601 User s Manual Michell Instruments 73 APPENDIX B Appendix B Hazardous Area Certification...
Page 86: ...QMA601 User s Manual 76 97449 Issue 3 June 2018 APPENDIX C Appendix C Modbus Register Map...
Page 111: ......
Page 112: ...http www michell com...