MDM300 I.S. User’s Manual
Michell Instruments
31
OPERATION
The display shows:
#
Indicates a log number for each set of log points
INT
Value of primary display parameter
EXT
Value of secondary display parameter
ST
Status column which reports the status of the instrument at each log
point in hexadecimal code.
The status codes are explained in Appendix B.
•
FILE
File name
•
INT
Primary display parameter being logged
•
EXT
Secondary display parameter being logged
•
INTVAL
Logging
interval
•
D#
Current date
in YYMMDD format
•
T#
Current time
in HHMMSS format
NOTE: The chart and datalogging are independent functions; therefore the
parameter displayed on the chart may not necessarily reflect what is being
logged.
3.4.11
CALIBRATION
!
WARNING: These procedures require the use of specialized
test equipment and calibration adjustments should only be
carried out by qualifi ed personnel.
POINT
CALIBRATION
-100
-90
-80
-70
-60
REF
-100.0
-90.0
-80.0
-70.0
-60.0
MDM300
-100.0
-90.0
-80.0
-70.0
-60.0
Figure 26
CALIBRATION Page
Entry to the CALIBRATION Menu is protected by a passcode (
4876
). The passcode is
entered in a similar manner to that described in Section 3.3.1.
This page gives access to the internal sensor’s calibration look-up table and provides the
facility to select and change any single point, or number of points.
The values in both the
REF
(reference dew-point) and
MDM300
I.S.
(reading at the
corresponding
REF
level input) columns can be edited.
Section 6 details the use of the calibration routines.
Summary of Contents for MDM300 I.S
Page 1: ...MDM300 I S Advanced Dew Point Hygrometer User s Manual I S 97213 Issue 5 October 2017...
Page 55: ...MDM300 I S User s Manual Michell Instruments 47 APPENDIX A Appendix A Technical Specifications...
Page 64: ...MDM300 I S User s Manual 56 97213 Issue 5 October 2017 APPENDIX D Appendix D FCC Declaration...
Page 70: ...MDM300 I S User s Manual 62 97213 Issue 5 October 2017 NOTES...
Page 71: ......
Page 72: ...http www michell com...