MDM300 I.S. User’s Manual
Michell Instruments
23
OPERATION
3.4.2
LOGGING
Logging can be set up either from the SETTINGS Menu or from the Chart Page (see
Section 3.4.8). Both methods of entry lead to the same LOGGING SET-UP menu as
shown below:
FILE NAME
PARAMETER
LOG INTERVAL
START?
-
DEWP
5
>
s
STOPPED
LOGGING
Figure 16
LOGGING Page
Parameter
Description
FILE NAME
De
fi
nes datalog
fi
le name. Field width up to eight alphanumeric
characters.
NOTE:
key deletes characters
NOTE: If a fi le name is not entered, a logging cycle cannot
be started.
Available Input:
0 - 9, A - Z
(keying up and down will change the
numbers to letters or vice-versa)
PARAMETER
Sets parameter to be logged. This need not necessarily be the same
as that displayed on the Main Display.
NOTE: Only one parameter may be logged at one time
Available Input:
DEWP, GM3, GKG, RH, PPMW, PPMV, CALC DP,
GM3(NG), LBMMSCF, PPMV(NG)
LOG INTERVAL
Sets log interval in 5 sec increments.
Available Input: Range
5 to 600 seconds
START ?
Starts the log cycle after logging parameters are set-up.
NOTE: If a fi le name is not entered, a logging cycle cannot
be started.
Once logging has started, log
fi
le parameter
fi
elds are locked out and
cannot be changed.
Available Input:
STOPPED, STARTED
Table 5
LOGGING Parameters
NOTE: If a duplicate file name has been entered an error message will be
shown when attempting to start logging.
A maximum of 64 log files can be stored in the instrument memory before
it is necessary to download or remove them. The
START ?
option will not be
available if this limit is exceeded.
Cycling power will stop the logging.
The size of any one log file is limited to 60kb. If the unit is left logging,
and the file size exceeds 60kb, then the log file will finish automatically. To
ensure that the log file you initiate collects all the data from your test, an
Summary of Contents for MDM300 I.S
Page 1: ...MDM300 I S Advanced Dew Point Hygrometer User s Manual I S 97213 Issue 5 October 2017...
Page 55: ...MDM300 I S User s Manual Michell Instruments 47 APPENDIX A Appendix A Technical Specifications...
Page 64: ...MDM300 I S User s Manual 56 97213 Issue 5 October 2017 APPENDIX D Appendix D FCC Declaration...
Page 70: ...MDM300 I S User s Manual 62 97213 Issue 5 October 2017 NOTES...
Page 71: ......
Page 72: ...http www michell com...