HygroSmart HS3 User’s Manual
14
97516 Issue 5, July 2017
SENSOR REPLACEMENT
Figure 18
Editing the Adjustment Table
After Correction:
Figure 19
Probe readings post-adjustment
2
nd
Adjustment Point
1.
Place the HS3-Probe together with a reference in a controlled and stable
environment of ± 50%RH.
2.
Check the readings of the HS3-Probe and the reference.
•
The Humidity reference is reading 50.2%RH.
•
The HS3-Probe reads 51.3%RH.