HygroSmart HS3 User’s Manual
Michell Instruments
15
SENSOR REPLACEMENT
Figure 20
Probe readings prior to adjustment
3.
Find the nearest point in the RH Adjustment table (in this example 50)
and change it into the measured reference value. (in this example 50.2).
4.
Subtract the Humidity value of the HS3-Probe from the reference value
(50.2 – 51.3 = -1.1).
5.
Add the outcome to the reference value (50.2 – 1.1 = 49.1).
6.
Enter this value in the corresponding
fi
eld of Corrected RH.
Figure 21
Editing the Adjustment Table