HygroSmart HS3 User’s Manual
12
97516 Issue 5, July 2017
OPERATION
To change a setting, either choose de
fi
ned values from the drop-down lists or enter
values into the text boxes under the
Change settings
column and then click the
Apply
button to apply the settings.
After applying new settings, the software automatically reads back the probe con
fi
guration
and displays it on screen in the left (current settings) column.
The
Advanced
button takes you to the Probe Adjustment window via a password entry
box. Enter the password
7316
to continue.
3.6 Probe
Adjustment
This window provides access to both the RH and temperature calibration tables, and the
probe Modbus address.
Picture below showing the default (factory) settings.
Figure 16
Probe Adjustment Screen Shot
To change a value, type within the text boxes provided and click the Apply... button to
apply the new value(s) to the HS3 Probe.
Important : the values in point 1 to 5 must be in ascending order!
Values are checked before they are applied to the HS3 Probe according to a set of rules.
The rules are displayed on the right hand side of the window. The software will display
a warning if any values do not comply with the rules.
Clicking the
Read
... button reads and display the tables from the HS3 Probe.
Clicking the
Save
... button saves the tables to a
fi
le.
Clicking the
Load
... button loads the previously saved
fi
le.xxxxxxx
RECOMMENDATION: It is a good idea to save the tables before editing them.
You can then Load (retrieve) the saved table in case any mistakes are made.
NOTE: If the Modbus address is changed, the software will continue
communicating with the probe under the new address.