![Michell Instruments Cermet II User Manual Download Page 37](http://html1.mh-extra.com/html/michell-instruments/cermet-ii/cermet-ii_user-manual_1784653037.webp)
Cermet II User’s Manual
30
97049 Issue 25.3, October 2016
APPENDIX B
Appendix B
Setup Codes
CAL
Calibration modes for input and output
DEFAULT VALUE = 052
Digit 1st digit (left most)
Calibration Mode
2nd digit
Calibration Function
3rd digit
Object for Calibration
0
Calibration functions as
per 2nd and 3rd digit
No function
No function
1
Set baud rate, parity and
serial address
Manual calibration (channel as per 3rd
digit)
Processed result (dew
point, ppm
V
, lbs/MMSCF,
g/m
3
)
2
N/A
N/A
Pressure input
3
N/A
N/A
N/A
4
N/A
N/A
N/A
5
N/A
Manual adjust
N/A
6
N/A
Analog output scaling (analog channel
as per 3rd digit)
N/A
CODE 1
Tendency Indication, Additional LED’s, Display data source, Flashing,
Decimal
points,
Rounding
DEFAULT VALUE = 200
Digit 1st digit (left most)
Additional LED’s
2nd digit
Display Data Source
3rd digit
0 only (No Function)
0
LED annunciators are
always off
Processed data - result (dew point,
ppm
V
, lbs/MMSCF, g/m
3
)
N/A
1
LED annunciators are
on when relays are de-
energized
Processed data - channel 1 (dew
point only)
N/A
2
LED annunciators are on
when relays are energized Processed data - pressure
N/A
CODE 3
Serial mode and analog output source
DEFAULT VALUE = 000
Digit 1st digit (left most)
Serial Mode
2nd digit
Analog Output 1 Source
3rd digit
0 only (No Function)
0
ASCII mode
Analog output 1 from processed
result data (dew point, ppm
V
, lbs/
MMSCF, g/m
3
)
N/A
1
N/A
Analog output 1 from processed data
(dew point only)
N/A
2
N/A
Analog output 1 from processed data
(pressure)
N/A