Service, maintenance & care

of the EDS unit model A

With the exception of the electromechanical valve,
mechanically operated switches and buttons, there
are no parts or sensors inside the EDS that will
age, cure or become out of tolerance in time and
use.  Therefore, EDS unit does not require any
regular or periodical maintenance or calibration.

If the unit is used in an application where dirt and dust are
present, the electromechanical valve may become clogged
with debris that may cause the flow to decrease or not allow it
to shut off completely. If this happens you can usually remedy
the problem by back-flushing the valve.  This is done by
supplying a small amount of air or oxygen pressure into the
large (6mm) outlet line while the electromechanical valve is
held in the full open state.

To clear a clogged valve follow these steps:

1

Remove the 6mm blue union from the outlet side of the

EDS unit.

2

Connect one end of the 4 mm tubing to the outlet of an

XCR regulator connected to an oxygen cylinder with at
least 50 psi of pressure and the other end into a 6mm to
4mm reducing union connected to the outlet of the EDS
unit to be flushed (purged)

If a reducing union is not available, connection can be

done by pressing (sliding) the 4mm tubing into the inside
diameter of the 6mm outlet tube on the EDS unit.

3

With the EDS unit on and the battery test button

depressed turn on the cylinder valve slowly to allow gas
to flow backwards through the EDS unit for about 10
seconds.  Then turn off the cylinder valve with the battery
test button still depressed.

If the above steps do not correct the described problem, the
EDS unit is not functioning correctly or if you would like it
checked our for calibration you can send it to Mtn. high E&S
co for diagnosis and repair.

While not being used the EDS unit should be stored in a secure
manner to ensure that dirt and debris do not enter and become
lodged in the inlet and outlet tubes.

The battery should be removed from the EDS unit if it is to be
stored for an extended period of time.

The electromechanical valve has a service life of well over
150 million pulsing cycles.  The respiration sensor has a
service life of well over 500 million pulsing cycles.  At this
time the valve and sensor may need replacing.  

Press battery test 
button while pressure 
is being applied to 
outlet tube to clear 
debris inside valve

Battery 
test 
button

6mm to 4mm 
reducing union

Outlet

Inlet

EDS unit

XCR 
regulator

Air or 
oxygen 
source

CAUTION

If the back-flush purging will not be done with an XCR

regulator do not apply pressures over 35 psig into the outlet

of the EDS without the battery test button being pressed, for

this may cause the breathing sensor to rupture.

Summary of Contents for EDS A-1

Page 1: ...erations up to 17 999 ft or a face mask for flight operations over 17 999 ft and up to 30 000 ft Both of these items are included with each EDS kit Several EDS units can share a single cylinder and pr...

Page 2: ...yed mode will arm the EDS unit to delay delivering oxygen until it is at a pressure altitude of 10 000 ft This is suitable for day time fight operations The power ON OFF switch makes connects and brea...

Page 3: ...e lapse rate curve that starts at sea level and lowers towards a partial pressure representing some very high altitude i e 100 000 ft The spa curve assumes that at sea level the barometric pressure is...

Page 4: ...ylinder The low pressure service line c is a high quality polyurethane line that is kink proof and flexible under varying temperatures The cannula and service line connects to the EDS unit via Quick C...

Page 5: ...ethane line without pushing in the collar It will damage the connector See figures below DO NOT ATTEMPT TO REMOVE REGULATOR FROM CYLINDER WHILE UNDER PRESSURE Doing so will destroy the O ring on the r...

Page 6: ...have their own nasal cannula and face mask In addition replacement of the nasal cannula should be done if it has become soiled or damaged in any way The flow fault alarm will respond to the following...

Page 7: ...the inside diameter of the 6mm outlet tube on the EDS unit 3 With the EDS unit on and the battery test button depressed turn on the cylinder valve slowly to allow gas to flow backwards through the ED...

Page 8: ...nto one standard cannula This allows the pilot to simply throw an on off switch to bypass the EDS unit then having the ability to adjust the flowmeter for the altitude to economize on oxygen usage Bec...

Page 9: ...unit The EDS EPS External Power Supply input jack and plug EDS EPS EDS unit 10 26 VDC input strip The EDS EPS with internal noise and glitch filter unit can operate up to six EDS units External Batter...

Page 10: ...C Nom 25 RH 25 C Max 100 RH 50 C near condensing Vibration Random vibration 5 to 500 Hz 15 minutes per axis 2 5 g rms sin wave Operating pressure range pressure altitude Min 3 4 psia 25 C appx 30 000...

Page 11: ...by Mt High E S Co at a nominal charge Before this warrantee can be valid Mountain High Equipment Supply Co must have your owners guarantee registration form properly filled out and in our files Mount...

Reviews: