4. Timer is running but the
meter reads 0 even when the
intensity is set at maximum.
Cycle the Main power control by
pressing it once and wait 3
seconds then press it again. This
Condition may be due to the
extended high current protection
described in Note # 2 below.
Note 1. The Sys*stim 206 is equipped with an interlock sensor that
makes it mandatory that the intensity be at the minimum output
(zero) when treatment begins and at any mode change. The
timer and treatment will not begin when Go is depressed unless
the Intensity control is at the Reset position (fully
counterclockwise).
Note 2. The Sys*stim 206 is equipped with extended high current
sensors that will shut down the output immediately upon an
error condition for the protection of the patient. The extended
high current condition may be reset by turning the Main power
control off for at least 3 seconds before turning it back on.
Caution: An extended high current condition that occurs when the patient
is attached and undergoing treatment constitutes a potentially
hazardous condition. If the condition repeats it self contact the
distributor or manufacturer. Do not continue to use the unit.
If problem is not addressed above, or if additional troubleshooting
guidance is desired, call (800) 854-9305 or email service directly at
.