treatment time in combination treatment. However stimulation
continues until hold button is depressed on the Sys*stim 206.
7. Select the desired current mode by depressing one of four selector
keys: 1) Narrow pulse; 2) Wide pulse; 3) AC; or 5) DC Continuous. (a
green indicator light will come on next to the mode key that is
selected.)
8. The surge feature is available for AC and pulse modes only. Depress
the control once and F35 is displayed. Press the waveform button and
F83 is displayed. These codes indicate the set frequency for the
selected Surge Mode.
9. Reverse polarity is available for all modes.
10. After completing steps 1-8, depress “Go” to initiate the selected
treatment. Note: Refer to the mode/indication table for the specific
output indication display for each mode and option combination.
11. Rotate the Intensity Control clockwise to increase output to the desired
treatment level. (Adjust intensity in Surge mode only when the green
Pulse Frequency indicator shows pulses present during the first 15
seconds of initial Surge mode operation. After that you will only be
able to adjust the intensity level up during the “On” time of the Surge
cycle.)
12. Rotate the Pulse Frequency Control clockwise to increase pulse
repetition to desired rate (green indicator next to this control flashes
with every output pulse.) Note: The Pulse Frequency Control is
defeated in DC Continuous and Surge modes.
13. Treatment can be stopped at any time by pressing the Hold switch.
The output will fall to zero and the Timer will stop but hold its time
remaining. Restart treatment by rotating the Intensity Control fully
counterclockwise to reset position then pressing the Go switch.
Proceed from step 11. Note: A two-second beeper activates at the
end of treatment time except in timer bypass mode.