Sys*Stim 206 Instruction Manual—
Rev. E_4/4/13
9
Section 2—Control Descriptions
and List of Abbreviations
2.1 Control
Descriptions
Mains power is controlled by an On/Off toggle switch. The
unit is off when the switch is pressed to the right and on
when pressed to the left.
00.00
Time display shows time in treatment in minutes and
seconds.
The timer keypad consists of eleven touch-type buttons
labeled 0-9 and timer bypass. Prior to starting an
operation, these buttons allow for entering the desired
treatment time from 1 to 29 minutes, or a continuous run
by depressing timer bypass. Either treatment time
remaining will be displayed on the treatment time
indicator, or elapsed time with flashing dashes when in
timer bypass mode. Bypass elapsed time is shown in
even seconds and restarts at 0:00 after each 29 minute
time period.
The “Go” button starts timer and begins the treatment by
providing the AC, pulse, or DC energy to the electrode
jack output if the intensity control is in fully
counterclockwise position (Reset) at the time the Go
switch is depressed.
The “Hold” button” halts the timer without disturbing the
remaining treatment time and stops the electrical output
when activated. The remaining treatment time will be
indicated.
Intensity is controlled by a three quarter turn adjustment
knob. This control varies the output voltage, or current, to
the patient during operation. Output increases with a
clockwise rotation, and decrease with a counterclockwise
rotation. Reset is achieved at the fully counterclockwise
position. Operation does not begin until the Intensity
control is returned to the Reset position.
This is the electrode output jack for plugging in electrode
cables. Use Mettler part number, ME 2260 to achieve best
results.