19
Sonicator 718 & 719 Palm-sized Ultrasound
Operating Procedures
8
Press the START/STOP switch
to initiate output.
* Any settings except for the program
settings can be edited during output. (See
pages 17 to 18.)
* If the ultrasound head is not in proper
contact with the body for approximately
3 minutes, the output will shut off
automatically.
* The audible therapy signal is turned on and
off from the Config. screen. (See page 22.)
* If no key operation occurs for approximately
3 minutes, power for the device will shut off
automatically.
P1
5
00
10
%
:
1.00
W / cm
2
LED ON
This is
displayed
during
output.
While ultrasound is being output,
as long as the ultrasound head is
in proper contact with the body,
an audible therapy signal will be
emitted and the output LED will
remain on.
Press the START/STOP switch to end output before the timer runs out.
* The timer will revert to the set time.
7
Apply the Ultrasound Gel
to the ultrasound head.