8
Sonicator 718 & 719 Palm-sized Ultrasound
To Ensure Correct and Safe Use
The following accessories for Sonicator 718 and Sonicator 719 are replaceable.
Call Mettler Electronics Corp. at 800-854-9305 to place an order.
7181
AC Adaptor (100–240 V)
7183
Power Cable for AC Adaptor
(110–120 V, only U.S.A & CANADA, Type B)
7188
Battery Case
7185
Battery Charger
7187
Power Cable for Battery Charger
(110–120 V, only U.S.A & CANADA, Type A)
7184
Rechargeable Battery
7191
Pouch
7189
Unit Stand
7180
Protective Cap for Sonicator 718 Probe
7190
Protective Cap for Sonicator 719 Probe
1844
Sonigel (12 X 9.5 oz. tubes)
1851
Sonigel (12 X 250 ml bottles)
1852
Sonigel (1 X 5 L)
1853
Sonigel (4 x 1852)
1863
Sonigel Lotion with Aloe Vera (1 G)
1864
Sonigel Lotion / Aloe Vera (4 X 1 G)
4. Maintenance and check items
Perform periodic inspections to ensure safety.
If you have any questions, contact the dealer or manufacturer or distributor.
Item
Description
Method
Appearance and
indications
Check the appearance for any signs of damage; whether
the display panel is distorted; and whether the indications
flicker. Make sure the ultrasound head is clean.
Check
visually.
Operation
Turn the POWER switch on and check that the device
operates normally and exhibits no problems.
Check by
operating.
Output
Place water on the ultrasound head and turn on the output.
Examine the water drops to assess the vibrations.
Check by
operating.
Ultrasound head
Check for cracks in the ultrasound head or problems with
the cable connection that may permit water or gel to enter
the device.
Check
visually.