Mettler Electronics Corp.—
Rev.F_07/03/12
20
4.3
User Program Memory
1. The Laser Sys*Stim 540 has ten program memories. The
practitioner can store commonly used settings and at a later time
recall them.
2.
Recalling User Program:
To recall a user program press the
PROGRAM key. “PROG 0” will be displayed and a short time later
the PROG 0 settings will be displayed. If the PROGRAM key is
pressed again the next program memory will be selected, which is
“PROG 1”. Following this procedure any of the (PROG 0 to PROG
9) memories can be selected by pressing the PROGRAM key until
the desired program memory is selected.
3. Once the desired program is selected, the settings can be used for
treatment by pressing the LASER ACTIVATE capacitance pad twice
(first press is to confirm, second press is for treatment as normal).
4. For example, using Program 2 for treatment.
Press PROGRAM key. “PROG 0” is displayed and a moment later
settings for PROG 0 are displayed.
Press PROGRAM key again. “PROG 1” is displayed and a
moment later settings for PROG 1 are displayed.
Press PROGRAM key. “PROG 2” is displayed and a moment later
settings for PROG 2 are displayed.
Press LASER ACTIVATE. Release LASER ACTIVATE (settings
are locked in)
Press LASER ACTIVATE >> PROG 2 treatment given.
5.
Storing User Program:
To store a user program in memory, the
user must select a program memory to modify. This is done by
pressing the PROGRAM key until the desired program memory is
selected. The program memory is modified by adjusting the Energy /
Time / Pulse Mode parameters as described in 4.2. To store these
new parameters, press and hold the PROGRAM key until two beeps
are heard.
6. For example, modifying and storing in Program 1 memory.
Press PROGRAM key. “PROG 0” is displayed and a moment later
settings for PROG 0 are displayed.
Press PROGRAM key again. “PROG 1” is displayed and a
moment later settings for PROG 1 are displayed.
Adjust Parameters.
Press and Hold PROGRAM key until two beeps are heard.
New Parameters are stored in PROG 1.