RM – Reverse Mains Test 

This test will break both the Live/L1 and Neutral/L2 power lines to the DUT. It will 
then reverse the normal connection, so that what was Live/L1 is now Neutral/L2, 
and what was Neutral/L2 is now Live/L1. To choose Reverse Mains measurements 
when the selection is available, press the 

rev. pol

 button. The display will reflect 

the change in measurement type. 
 

 

Warning! 

When testing units that have a large inductive load (motor), 

always turn OFF the DUT 

BEFORE

 reversing Mains. 

 

Mains Voltage Test – “V”  

This test measures the mains voltage, in volts, 
between Live and Neutral. Press the 

button below 

the 

line voltage

 label, and the result will be 

displayed. 

 

Protective Earth Test – “

m

” 

This test measures the resistance, in milliohms, 
between the enclosure of the DUT and Earth contact 
on the front outlet. To execute the test and display 
the results, connect the test lead to the enclosure of the DUT, and then press the 

m

 button below the 

gnd wire res

 label.  

 

Earth Leakage Current Test – “

µ

A” – ON, RM 

This test measures the current, in microamps, 
passing through the Earth lead. To execute the test 
and display the results, press the 

earth

 button below 

the 

leakage current

 label. 

 

 

Warning! 

When testing units that have a large inductive load (motor), 

always turn OFF the DUT 

BEFORE

 reversing Mains. 

 

Enclosure Leakage Current Test – “

µ

A” – OE, 

ON, RM, OERM, ONRM 

This test measures the leakage current, in 
microamps, between the enclosure of the DUT and 
Earth contact on the front outlet. To execute the test and display the results, connect 
the test lead to the enclosure of the DUT, and then press the 

enclosure

 button 

below the 

leakage current

 label. 

 

 

Warning! 

When testing units that have a large inductive load (motor), 

always turn OFF the DUT 

BEFORE

 reversing Mains. 

Protective Earth 

130m

 

Enclosure Leakage 

Current 

xxx

µ

Earth Leakage Current 

xxx

µ

Mains: NC 

Mains Voltage 

120V 

Summary of Contents for QA-ST MKII

Page 1: ...QA ST MKII Electrical Safety Tester US Model Getting Started Guide...

Page 2: ...ially conform to published specifications and to the documentation provided that it is used for the purpose for which it was designed METRON will for a period of twelve 12 months from date of purchase...

Page 3: ...rement Selection 2 OE Open Earth Test 2 ON Open Neutral Test 2 RM Reverse Mains Test 3 Mains Voltage Test 3 Protective Earth Test 3 Earth Leakage Current Test 3 Enclosure Leakage Current Test 3 Patien...

Page 4: ...t with power 2 Confirm that the only lights illuminated are the two lights in the inlet power indicator boxed in blue If either of these lights is not illuminated or if any other combination of lights...

Page 5: ...ovided test name unit of measure and options Options include assorted user selectable power conditions as well as the AC DC measurement selection AC DC Measurement Selection To select AC or DC measure...

Page 6: ...nt outlet To execute the test and display the results connect the test lead to the enclosure of the DUT and then press the m button below the gnd wire res label Earth Leakage Current Test A ON RM This...

Page 7: ...ient lead to all other patient leads in parallel To select and display the leakage of an individual patient lead to all other leads in parallel repeatedly press the lead to lead button below the patie...

Page 8: ...LEAKAGE CURRENT MEASUREMENTS DC AND AC Range 0 1999 A Accuracy DC and 25 to 1 kHz 2 of reading 3 A Resolution 1 A Test modes AC and DC measurements LEAD ISO MEASUREMENTS Range 0 6000 A RMS Accuracy D...

Page 9: ...462722 Accessories 11042 User Manual CD Part Number 11042 11150 Carrying Case Part Number 2462192 17260 Hard Case Part Number 2462161 17358 Test lead Kelvin cable Part Number 2568891 17024 Universal b...

Reviews: