METRON QA-ST MKII Getting Started Manual Download Page 5

QA-ST MKII – 

Getting Started Guide

 

 

 

2

Setup for Automated Testing 

 

1.

 

Perform all steps under “Setup for Manual Testing” on the previous page. 
 

2.

 

Connect one end of the NULL-modem cable to the RS-232 port on the left side 
of the 

QA-ST MKII

.  

 

3.

 

Connect the other end of the NULL-modem cable to an applicable COM port on 
your computer. 
 

4.

 

Consult the 

ansur test executive User Manual

 and the 

ansur QA-ST Plug-In User 

Manual

 for instructions on building and executing the automated test templates. 

 

5.

 

Perform any of the 

QA-ST MKII

 electrical safety tests. 

 
 

4 Tests 

 

For each test listed below, the following information is provided: test name, unit of 
measure, and options. (“Options” include assorted, user-selectable power conditions, 
as well as the AC/DC measurement selection.) 
 
 

AC/DC Measurement Selection 

To select AC or DC measurements when the selection is available, press the 

AC/DC

 

button. The display will reflect the change in measurement type. 

 

OE – Open Earth Test 

This test will break the connection of the Earth power line to the DUT. To select 
Open Earth measurements when the selection is available, press the 

open gnd

 

button. The display will reflect the change in measurement type. 

 

ON – Open Neutral Test  

This test will break the connection of the Neutral/L2 power line to the DUT. To 
select Open Neutral measurements when the selection is available, press the 

open 

neutral

 button. The display will reflect the change in measurement type. 

 
 

Summary of Contents for QA-ST MKII

Page 1: ...QA ST MKII Electrical Safety Tester US Model Getting Started Guide...

Page 2: ...ially conform to published specifications and to the documentation provided that it is used for the purpose for which it was designed METRON will for a period of twelve 12 months from date of purchase...

Page 3: ...rement Selection 2 OE Open Earth Test 2 ON Open Neutral Test 2 RM Reverse Mains Test 3 Mains Voltage Test 3 Protective Earth Test 3 Earth Leakage Current Test 3 Enclosure Leakage Current Test 3 Patien...

Page 4: ...t with power 2 Confirm that the only lights illuminated are the two lights in the inlet power indicator boxed in blue If either of these lights is not illuminated or if any other combination of lights...

Page 5: ...ovided test name unit of measure and options Options include assorted user selectable power conditions as well as the AC DC measurement selection AC DC Measurement Selection To select AC or DC measure...

Page 6: ...nt outlet To execute the test and display the results connect the test lead to the enclosure of the DUT and then press the m button below the gnd wire res label Earth Leakage Current Test A ON RM This...

Page 7: ...ient lead to all other patient leads in parallel To select and display the leakage of an individual patient lead to all other leads in parallel repeatedly press the lead to lead button below the patie...

Page 8: ...LEAKAGE CURRENT MEASUREMENTS DC AND AC Range 0 1999 A Accuracy DC and 25 to 1 kHz 2 of reading 3 A Resolution 1 A Test modes AC and DC measurements LEAD ISO MEASUREMENTS Range 0 6000 A RMS Accuracy D...

Page 9: ...462722 Accessories 11042 User Manual CD Part Number 11042 11150 Carrying Case Part Number 2462192 17260 Hard Case Part Number 2462161 17358 Test lead Kelvin cable Part Number 2568891 17024 Universal b...

Reviews: