QA-ST MKII –
Getting Started Guide
1
1
Terms and Abbreviations
Chassis = Enclosure
DUT = Device Under Test
Earth = Ground
Enclosure = Chassis
Ground = Earth
NC = Normal Condition
OE = Open Earth
ON = Open Neutral
RM = Reverse Mains
This symbol indicates that mains voltage or higher will be present in
the test leads. In addition, the enclosure may be raised to mains voltage
levels or higher.
2
Setup for Manual Testing
1.
Plug the unit into a wall outlet with power.
2.
Confirm that the
only
lights illuminated are the two lights in the
inlet power
indicator (boxed in blue). If either of these lights is not illuminated, or if any
other combination of lights is illuminated, examine the wall outlet for problems
in grounding or polarity.
NOTE:
DO
NOT
USE extension cords or power strips when testing with the
QA-ST MKII
.
3.
Connect the test cable to the
Enclosure
port on the right side of the unit.
4.
Connect the test cable clamp to the metal enclosure of your DUT.
5.
Connect the patient leads to the
Patient leads
ports in numerical order.
6.
Perform any of the
QA-ST MKII
electrical safety tests.