2600 - 5200 ohms in steps of 100 ohms (@
dc).
Additional Fixed Load:
200 ohms, 500 watt maximum.
Crest Factor:
The higher of the two peak-measurements is
used for calculation.
Range:
1.4 - 16 (V peak voltage / V RMS).
Foot Switch Output:
The output triggers the measurement after a
programmed delay time, defined as the time
period from the activation of the foot switch to
the beginning of data processing. The delay
time is 200 ms - 4000 ms.
Peak To Peak Voltage:
0 to 10 kV (closed load only) ± 10%. Mea-
surement is taken between the active and
dispersive electrodes with closed load only.
Oscilloscope Output:
5 V/A uncalibrated, 100 mA RF current mini-
mum input.
Isolation:
10 kV isolation between measurement device
and enclosure.
1.3 General Information
Temperature Requirements:
+15
°
C to +35
°
C when operating
0
°
C to +50
°
C in storage
Display:
Type
LCD graphic display
Alphanumeric format
8 lines, 40 characters
Graphics mode:
240 x 64 point matrix
Display control:
5 F-keys, enter, cancel and an encoder
Data Input/ Output (2):
Parallel printer port (1); Bi-directional RS
-232C (1) for Computer control
Power Source:
From 115 VAC to 230 VAC, 48/66 MHz.
Mechanical Specifications:
Housing
Metal case
Height
13.2 cm / 4.48 in.
Width
34.2 cm / 11.61 in.
Length
39.5 cm / 13.41 in.
Weight
9.8 kg / 21.6 lbs.
Standard Accessories:
QA-ES Electrosurgical Analyzer
(P.N. 14010)
Power Cord
(P.N. 14300)
QA-ES User and Service Manual
(P.N. 14025)
Additional Accessories:
E-Input Measuring Cable - Black
(P.N. 11451)
2
Summary of Contents for QA-ES
Page 1: ...QA ES User Service Manual QA ES Electrosurgical Analyzer P N 14025...
Page 4: ...This page intentionally left blank iv...
Page 6: ...This page intentionally left blank vi...
Page 10: ...This page intentionally left blank 4...
Page 14: ...This page intentionally left blank 4...
Page 24: ...This page intentionally left blank 4...
Page 40: ...This page intentionally left blank 12...