Test Configuration
Limits of Acceptable Leakage Current
Bipolar
The leakage current should not exceed 1% of the
maximum bipolar rated power output.
Measured at equipment
terminals
The leakage current should not exceed 100 mA.
1. Grounded HF Equipment: Measurements of the HF cur-
rent leakage.
The ESU is grounded. The test load is 200 ohms
and the ESU must be operating at maximum power. The current
leakage measured directly at the instrument's terminals must not
exceed 100 mA.
Test setup in compliance with IEC 601.2.2, sec. 19.101a, test 1,
fig. 102 and sec. 19.102. (Adopted by ANSI/AAMI HF18-
1993)
Test setup in compliance with IEC 601.2.2, sec. 19.101a, test 2,
fig. 103 and sec. 19.102. (Adopted by ANSI/AAMI HF18-1993)
2. HF Isolated Equipment: Measurements of the HF current
leakage from the active and neutral electrodes.
The test load
is 200 ohm and the ESU must be operating at maximum power.
The current leakage measured directly at the instrument's termi-
nals must not exceed 100 mA.
Active electrode test setup in compliance with IEC 601.2.2, sec.
19.101b, fig, 104 and sec. 19.102. (Adopted by ANSI/AAMI
HF18-1993)