100 ms from 1000 ms to 4000 ms.
Save the chosen delay in ‘Delay’ under the STATUS field by
pressing
ENTER (F5)
. Press
CANCEL (F4)
to undo.
5.
START (F3).
When you press on
START
, the test procedure
will begin, and the text in the field ‘Oper.’ will change from
‘Ready’ to ‘Measuring’. If the unit is set to the position for a
REM test, this text will change from ‘Ready’ to ‘lncr'. res.’
Press
STOP (F3)
to stop the test procedure.
6.
SETUP (F4).
Here you can set the power distribution level for
start, stop and step in ohms.
Choose the ‘Start load’ by using
KNOB PARAM (F4)
. (see
stars). Use the encoder to set the level. Save the level by press-
ing
ENTER (F5)
. Press
CANCEL (F4)
to undo. Go to ‘End
load’ and ‘Step Size’ and repeat the same procedure.
Pwr. distr. Start load
is the first load to be used during the
measurements; it can be set from 10 ohms to 2100 ohms, with
steps of 25 ohms starting at 25 ohms onwards.
Pwr. distr. End load
is the last load used in the measurements;
R can be set from 525 ohms to 5200 ohms, with steps of 25
ohms from 525 to 2500 ohms and step of 100 ohms from 2500
ohms to 5200 ohms
Pwr, distr. Step Size
is the load set with steps of 25, 50, 100,
and 200 ohms
Press
QUIT MENU (F5)
to return to the main menu.
7.
PRINT HEADER (F3).
Writes a heading for a new test proto-
col.
8.
REMOTE CONTR. (F4).
Enables you to control the QA-ES
through a PC. Required software: PRO-Soft QA-ES.
5
Summary of Contents for QA-ES
Page 1: ...QA ES User Service Manual QA ES Electrosurgical Analyzer P N 14025...
Page 4: ...This page intentionally left blank iv...
Page 6: ...This page intentionally left blank vi...
Page 10: ...This page intentionally left blank 4...
Page 14: ...This page intentionally left blank 4...
Page 24: ...This page intentionally left blank 4...
Page 40: ...This page intentionally left blank 12...