on the
QA-ES
. Don't activate any functions. Measure the cur-
rent. Required value:
230±10%: 80mA -10/+20mA
115±10%: 160mA -20/+40mA
2.
User Interface and Display.
Check that the display and the
user interface are working normally.
3.
Cooling Fan.
Check that the fan increases speed when you
press on
Start (F3)
in the main menu
and that the speed de-
creases about 10 seconds after you press
STOP (F3)
.
4.
Serial and Parallel Interface.
QA-ES can be connected to a
printer and a PC. Check that booth the serial and parallel inter-
faces work.
5.
Foot Switch.
Connect a multimeter to the foot switch output
on the QA-ES (green). Check that the relay for the foot switch
is connected when you press
START (F3)
.
6.
Fixed Load.
Measure the value of the fixed load with a multi-
meter (blue). Required value: 200 ohms ±10 ohms.
7.
Housing
Isolation.
Use a multimeter to check the isolation be-
tween the housing and the measuring inputs. It is important to
check all of the terminals on the housing: ‘Var. Load', 'Fixed
Load' and 'Foot Switch’.
5.4 Calibration
1.
Offset Voltage
.
Short-circuit the RF input on the sampling unit
(J4). All the measurements and reference values in this section
refer to the sampling unit. Measure the voltage at each of the
following points and adjust them to the following values with
the potentiometers specified between brackets. The voltage
should be set as close to 0V as possible.
TP1 - GND
(no adjust)
TP3 - TP2
(R3)
TP4 - GND
(R2)
TP5 - GND
(R1)
TP6 - GND
(R4)
OmV
∀
100
Φ
V
O
Φ
V
∀
50
Φ
V
O
Φ
V
∀
100
Φ
V
O
Φ
V
∀
500
Φ
V
OmV
∀
5 mV
2.
Load Resistance
. Set the QA-ES in calibration mode by hold-
ing the function key
F1
while switching the instrument ON.
Continue to press on the key until the main menu appears. Con-
nect a multimeter (hp 34401A) to the 'Var. Load' input. The re-
sistances are calibrated by turning the small wheel until the dis-
play shows a value as close as possible to the value read. To
calibrate the next resistance, press on
UP (F1)
or
DOWN (F2)
.
3.
Measuring Resistance
. Set the QA-ES in calibration mode.
Before mounting it, the measuring resistance R1 on the Load
Board should be measured with hp 34401A. The measured val-
ue should be noted near R1 on the Load Board. Use this value
for R1 in the following equation and for ‘Meas. Reas.’ in the
calibration menu by using the wheel. Press
QUIT (F5)
to save
calibrated values.
It is important to cali-
brate the resistance of
the measuring leads on
the multimeter itself
be-
fore
starting the test.
2
Summary of Contents for QA-ES
Page 1: ...QA ES User Service Manual QA ES Electrosurgical Analyzer P N 14025...
Page 4: ...This page intentionally left blank iv...
Page 6: ...This page intentionally left blank vi...
Page 10: ...This page intentionally left blank 4...
Page 14: ...This page intentionally left blank 4...
Page 24: ...This page intentionally left blank 4...
Page 40: ...This page intentionally left blank 12...