5-14
b.
Select
Automatic
in
Test Mode (F5)
. Press
START (F4)
.
QA-90 will then commence an automated sequence test.
c.
When the test is completed the following Test Results Win-
dow appears.
---TEST FAILED--- F7
Print Test Results ....................> F6
Print Failed Results ..................> F5
SHOW TEST SHOW FAILED STORE MAIN MENU
F1 F2 F3 F4
d.
Press
Print test results (F6)
. Check the printout for legibili-
ty. Then, press
Print failed results (F5)
. Check the printout
for legibility.
5.5 Calibration Constants
1.
SP/VMDL U-constant
. Adjusts the voltage measurement for
Supply Voltage and Voltage Measurement Dual Lead. When the
parameter value increases, the measurement value decreases. Ca-
librates at 200 - 230 V alternating voltage. This is best done at
Dual Lead. Measure with AC voltage with polarity changed. Add
the measurements to get the average and calculate the correct
constant.
2.
PEIRMDL U-constant
. Adjust the voltage measurement for
Protective Earth and Resistance Measurement Dual Lead. When
the parameter value increases, the measurement value decreases.
This value does not usually change. The default value is 84.30.
3.
IR U-constant
. Adjust the resistance measurement for insulating
resistance. When the parameter value increases, the measurement
value increases. The calibration occurs with a calibrating resis-
tance of 90 - 110 m
Ω
. Ensure the power of 500 V is set correctly,
and that the QA-90 has reached its operating temperature.
4.
General I-constant
. Adjust the current measurement for PLC,
ALC, MAP, EnLC, ELC, CMDL and the IR tests. When the pa-
rameter value increases, the measurement value decreases. The
IMPORTANT
ENSURE THAT YOU SAVE THE CALIBRATION CONSTANTS
TO FLASH.
Summary of Contents for QA-90
Page 1: ...QA 90 User Service Manual QA 90 Electrical Safety Analyzer P N 11025 V 3 13...
Page 14: ...2 4 This page intentionally left blank...
Page 30: ...3 16 This page intentionally left blank...
Page 40: ...4 10 MANUAL TEST SETUP F7 Mains Voltage F6 Current Consumption F5 MORE MAIN MENU F1 F2 F3 F4...
Page 44: ...4 14 This page intentionally left blank...
Page 51: ...5 7 MORE GO BACK MAIN MENU F1 F2 F3 F4...
Page 60: ......
Page 68: ...6 8 This page intentionally left blank...
Page 86: ...A 18 This page intentionally left blank...
Page 88: ...B 2 This page intentionally left blank...
Page 90: ...B 4 Measuring Board Component Location Layer 1...
Page 91: ...B 5 Measuring Board Component Location Layer 2...
Page 92: ...B 6 Measuring System Schematic Diagram Part 1 Measuring Preparation...
Page 93: ...B 7 Measuring System Schematic Diagram Part 2 High Voltage Interface...
Page 94: ...B 8 Measuring System Schematic Diagram Part 3 High Voltage Logic...
Page 97: ...B 11 Measuring System Schematic Diagram Part 6 Measurement Matrix...
Page 98: ...B 12 Measuring System Schematic Diagram Part 7 Program Amplifier and Lowpass...
Page 99: ...B 13 Measuring System Schematic Diagram Part 8 Relay Drivers...
Page 100: ...B 14 Microprocessor Board Component Location...
Page 102: ...B 16 Microprocessor System Schematic Diagram Part 1 QA 90...
Page 103: ...B 17 Microprocessor System Schematic Diagram Part 2 Integrated Keypad...
Page 104: ...B 18 Microprocessor System Schematic Diagram Part 3 CPU...
Page 105: ...B 19 Microprocessor System Schematic Diagram Part 4 CPU...
Page 106: ...B 20 Microprocessor System Schematic Diagram Part 5 Printer and Display Interface...
Page 108: ...B 22 This page intentionally left blank...
Page 110: ...C 2 This page intentionally left blank...
Page 112: ...D 2...