5-11
g.
Using a reference resistor of 90-110 M
Ω
, calculate the IR U-
constant
h.
Using a reference resistor of 190-210 M
Ω
, calculate the IR
R-constant such that the QA-90 measures as closely as poss-
ible to that reference.
such that the QA-90 measures as closely as possi-
ble to that reference.
i.
If you are unable to justify levels above 210 M
Ω
, it may be
necessary to change the measurement amplifier.
j.
Check the range of 1-200 M
Ω
(e.g. 1, 10, 100 and 200 M
Ω
)
with Metron's test instrument. Press
STOP (F1)
.
k.
Press
GO BACK (F2)
to return to Main Setup Window 2
and press
Mains on Applied Parts (F6)
.
l.
Select
(F7)
to display
Applied Part to Case
.
m.
Define the Patient Leads 1-10.
n.
Measure the voltage between EARTH and applied part.
Adjust the voltage to
253 V
AC
for European Models or
132 V
AC
for US Models
(Patient Leads 1-10).
o.
Check that the current displayed on the UUT is in accor-
dance with Ohm’s Law (253 or 132 V
AC
/test-box imped-
ance). The measured alternating voltage, divided by the
voltmeter’s internal resistance.
p.
Press
GO BACK (F2)
to return to the previous display.
5.4 Other Calibration Pro-
cedures
1.
Bar Code Reader
(Optional). This checks the QA-90 Bar Code
Reader function and its ability to properly accept scan and con-
2
On older QA-90 firmware versions there is only one constant IR U-constant
for insulation resistance measurements. On version series 0.xx there are six
constants, while on version series 1.xx there are seven. Also with the 1.xx
series, the IR U- and IR R-constants are used for insulation resistance mea-
surements.
NOTE
Remember that the patient
leads must be defined (1-10).
IMPORTANT
YOU HAVE TO PERFORM SEVERAL CALCULATIONS OF
THESE CONSTANTS BECAUSE OF THEIR INFLUENCE ON
EACH OTHER. IF THEY ARE CLOSE TO DEFAULT, FEW CAL-
CULATIONS ARE NECESSARY. HOWEVER, IF THE IR-R
CONSTANT HAS TO BE HIGH, MEASUREMENTS MUST BE
REPEATED TO OBTAIN THE BEST POSSIBLE LINEARITY.
NOTE
Never set any of the con-
stants higher than 99.999.
Summary of Contents for QA-90
Page 1: ...QA 90 User Service Manual QA 90 Electrical Safety Analyzer P N 11025 V 3 13...
Page 14: ...2 4 This page intentionally left blank...
Page 30: ...3 16 This page intentionally left blank...
Page 40: ...4 10 MANUAL TEST SETUP F7 Mains Voltage F6 Current Consumption F5 MORE MAIN MENU F1 F2 F3 F4...
Page 44: ...4 14 This page intentionally left blank...
Page 51: ...5 7 MORE GO BACK MAIN MENU F1 F2 F3 F4...
Page 60: ......
Page 68: ...6 8 This page intentionally left blank...
Page 86: ...A 18 This page intentionally left blank...
Page 88: ...B 2 This page intentionally left blank...
Page 90: ...B 4 Measuring Board Component Location Layer 1...
Page 91: ...B 5 Measuring Board Component Location Layer 2...
Page 92: ...B 6 Measuring System Schematic Diagram Part 1 Measuring Preparation...
Page 93: ...B 7 Measuring System Schematic Diagram Part 2 High Voltage Interface...
Page 94: ...B 8 Measuring System Schematic Diagram Part 3 High Voltage Logic...
Page 97: ...B 11 Measuring System Schematic Diagram Part 6 Measurement Matrix...
Page 98: ...B 12 Measuring System Schematic Diagram Part 7 Program Amplifier and Lowpass...
Page 99: ...B 13 Measuring System Schematic Diagram Part 8 Relay Drivers...
Page 100: ...B 14 Microprocessor Board Component Location...
Page 102: ...B 16 Microprocessor System Schematic Diagram Part 1 QA 90...
Page 103: ...B 17 Microprocessor System Schematic Diagram Part 2 Integrated Keypad...
Page 104: ...B 18 Microprocessor System Schematic Diagram Part 3 CPU...
Page 105: ...B 19 Microprocessor System Schematic Diagram Part 4 CPU...
Page 106: ...B 20 Microprocessor System Schematic Diagram Part 5 Printer and Display Interface...
Page 108: ...B 22 This page intentionally left blank...
Page 110: ...C 2 This page intentionally left blank...
Page 112: ...D 2...