METRON daeg SPO2 User Manual Download Page 14

 

 

 

www.metron-biomed.com

 

14 

 

Function Selection Menu 

 
This screen will let you select which testing function of 
the daeg that you wish to perform.  Using the 

Up

 and 

Down

 arrows will let you move the cursor (

->

) next to the 

function that you wish to use.  Once the cursor is placed 
press the (

F2

) button or the (

yes/enter

) button to enter 

that function.  Pressing the (

F3

) or (

esc

) buttons will take 

you back to the 

Main Simulation Screen

 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 

 

Alarm Limits 

 
This test is designed to test your alarm system on your 
monitor.  The (

F2

) button will move the cursor, (

*

), to the 

next parameter.  Using the Up and Down arrows will let 
you change the value of the selected parameter.  You will 
then need to press the (

yes/enter

) button to accept the 

change and start the timer.  Press the (

F1

) button to stop 

the clock when you hear the alarm from your monitor.  
The (

F3

) button will generate a line of printout via the 

parallel port.  Pressing the (

esc

) button will take you back 

to the 

Main Simulation Screen

 

 
 
 
 
 
 
 

-----FUNCTIONS------ 
  Alarm limits 
  Print header 
  Print Status 
  Remote Contr 

 

 
       ENTER    EXIT

 

F3 

F2 

F1 

-----FUNCTIONS------ 
  Print Status 
  Remote Contr 
  Probe Analyz 
  Sim. Setup 
 
       ENTER    EXIT

 

F3 

F2 

F1 

AL_LIMIT t= 0.0 sec 
Pulse :*On 
SpO2  :  96  % 
Rate  :  60  BPM 
P.ampl:  100 % 
 
STOPt   

*

 

    PRINT

 

F3 

F2 

F1 

Summary of Contents for daeg SPO2

Page 1: ...daeg SPO2 Simulator User Manual Version 1 15_2 Updated 22 01 2007...

Page 2: ...SICAL DAMAGE CAUSED BY MISUSE OR PHYSICAL ABUSE IS NOT COVERED UNDER THE WARRANTY Items such as cables and nonserialized modules are not covered under this warranty Recalibration of instruments is not...

Page 3: ...e g cables carrying cases auxiliary modules etc are not eligible for return or refund Only products returned within 90 days from the date of original purchase are eligible for refund credit In order...

Page 4: ...Calibration measurements are traceable to the National Institute of Standards and Technology NIST Devices for which there are no NIST calibration standards are measured against in house performance st...

Page 5: ...Table of Contents daeg Specifications 7 daeg Quick Start 9 Unit Setup 11 Operation 13 Troubleshooting 19...

Page 6: ...www metron biomed com 6 This page was intentionally left blank...

Page 7: ...curacy 2 1mV System rate accuracy 0 1 System time resolution 10 s 100 kHz maximum update rate ECG OUTPUT The data below is related to factory installed waveform types NORMAL ECG SINUS 30 300 BPM resol...

Page 8: ...www metron biomed com 8 This page was intentionally left blank...

Page 9: ...djust the active parameter using the yellow Up and Down arrow keys on the keypad then press enter to confirm the change 3 To change the simulation setting for oximeter model Make Ambient Light Pigment...

Page 10: ...www metron biomed com 10 This page was intentionally left blank...

Page 11: ...Refer to the diagram below when installing the batteries You will need to depress the tabs shown in the diagram below and remove the bottom dark gray piece of the chassis This will then expose the bat...

Page 12: ...www metron biomed com 12 This page was intentionally left blank...

Page 13: ...ing the F3 button The F2 button will move the cursor to the next parameter Using the Up and Down arrows will let you change the value of the selected parameter You will then need to press the yes ente...

Page 14: ...e F2 button will move the cursor to the next parameter Using the Up and Down arrows will let you change the value of the selected parameter You will then need to press the yes enter button to accept t...

Page 15: ...elect which test you wish to perform on the finger probe when you have the probe tester adaptor box attached Diode Test Pressing DIODE runs the test This test gives a descriptive message as to the con...

Page 16: ...ort in that element A Middle line is normal operation A Low indicates an open circuit Since this is a continuously running test it lets you flex you cable to check for intermittent failures Sensitivit...

Page 17: ...as follows Up and Down arrows control the cursor position yes enter increases the indicated value no clear decreases the selected value F1 decreases the selected SPO2 value F2 increases the selected S...

Page 18: ...www metron biomed com 18 This page was intentionally left blank...

Page 19: ...not match A Confirm that the Make is properly selected Confirm your selection by pressing the ENTER key after changing the value Q The monitor is not reading any values A Check all connections and ve...

Page 20: ...www metron biomed com 20 This page was intentionally left blank...

Reviews: