www.metron-biomed.com
13
Operation
Power up Screen
This screen is the first one that is visible when the unit is
turned on. It indicates which version of the firmware is
installed.
Main Simulation Screens
These are the two screens that are used when performing
SPO2 simulations. You can change between the screens by
pressing the (
F3
) button. The (
F2
) button will move the
cursor, (
*
), to the next parameter. Using the Up and Down
arrows will let you change the value of the selected
parameter. You will then need to press the (
yes/enter
)
button to accept the change. Pressing the (
F1
) button will
take you the
Function Selection Menu
.
---- METRON daeg ---
SpO2 Meter Analyzer
Revision X.XX
loading sim data...
F3
F2
F1
------ STATUS ------
Preset:*Manual
SpO2 : 96 %
Rate : 60 BPM
P.ampl: 100 %
FUNC.
*
displ-1
F3
F2
F1
------ SETUP -------
Make :*Nellcor
Oper. : Manual
Amb.L.: Manual
Pigm. : Manual
FUNC.
*
displ-2
F3
F2
F1