3.2. Remote control commands
719 S Titrino
84
$G sets all RS settings. The changes are performed only if the instrument is
inactive. After the setting of the interface parameters, wait at least 2 s to allow
the components to equilibrate.
Settings of the values for the data transmission via the RS interface: baud
rate, data bit, stop bit, parity and type of handshake, see also page 95ff.
The setting of the values must be initiated with $G immediately after entry of
the values.
3.2.2.50. C
onfig
.C
omVar.
C30
with up to
.C39
, etc.
0... +999 999
Values of the common variables from C30 up to C39. Insert the common
variables directly or describe the determination results directly from the
method, see 3.2.2.6.
3.2.2.51. S
mplData
.S
tatus ON,
OFF
On/off switching of silo memory. When the silo memory is switched on, the
sample data are fetched from the lowest valid silo line.
3.2.2.52. S
mplData
.O
FFSilo
.I
d1
up to 8 ASCII characters
S
mplData
.O
FFSilo
.Id2
up to 8 ASCII characters
S
mplData
.O
FFSilo
.Id3
up to 8 ASCII characters
S
mplData
.O
FFSilo
.V
alSmpl
6-digits, sign and decimal point
S
mplData
.O
FFSilo
.U
nitSmpl
up to 5 ASCII characters
Current sample data.
The identifications Id1...Id3 can be used in formulas as sample-specific calcu-
lation constants C21...C23.
If "no unit" is desired for the unit of the sample size, the blank string must be
entered.
3.2.2.53. S
mplData
.ON
Silo
.C
ounter
.M
axLines read
only
S
mplData
.ON
Silo
.C
ounter
.F
irstLine read
only
S
mplData
.ON
Silo
.C
ounter
.L
astLine read
only
Information on silo memory.
.MaxLines: Maximum possible number of silo lines.
.FirstLine: Lowest valid silo line.
.LastLine: Last occupied silo line.
3.2.2.54. S
mplData
.ON
Silo
.E
ditLine
.1.M
ethod
up to 8 ASCII characters
S
mplData
.ON
Silo
.E
ditLine
.1.I
d1
up to 8 ASCII characters
S
mplData
.ON
Silo
.E
ditLine
.1.Id2
up to 8 ASCII characters
S
mplData
.ON
Silo
.E
ditLine
.1.Id3
up to 8 ASCII characters
S
mplData
.ON
Silo
.E
ditLine
.1.V
alSmpl 6-digits, sign and dec.point
S
mplData
.ON
Silo
.E
ditLine
.1.U
nitSmpl
up to 5 ASCII characters
etc., up to
.99
Contents of a silo line.
.Method: Method used to process the sample, from the method memory or
from the card.
.Id:
The identifications Id1...Id3 can also be used as sample-specific
calculation constants C21...C23 in formulas.