![Metrohm Titrino 719 S Instructions For Use Manual Download Page 61](http://html1.mh-extra.com/html/metrohm/titrino-719-s/titrino-719-s_instructions-for-use-manual_1779645061.webp)
3.1 General rules
719 S Titrino
57
3.1.2 Triggers
Triggers initiate an action on the Titrino, for example, starting a process or
sending data. Triggers are marked by the introductory symbol $.
The following triggers are possible:
$G
G
o
Starts processes, for ex. starting the mode run or setting the RS 232 interface
parameters
$S S
top Stops
processes
$H H
old Holds
processes
$C C
ontinue
Continues processes after Hold
$Q Q
uery
Queries all information from the current node in the tree forward up to and
including the values
$Q.P P
ath
Queries the path from the root of the tree up to the current node
$Q.H H
ighest
Queries the number of son nodes of the current node
Index
$Q.N"i"
N
ame
Queries the name of the son node with index i, i = 1 - n
$D D
etail-Info Queries the detailed status information
$U
q
U
it
Aborts the data flow of the instrument, for example, after $Q
The triggers $G and $S are linked to particular objects, see the summary table
page 63ff.
All other triggers can be used at any time and at all locations on the object
tree.
Examples:
Querying the value of the baud rate:
&C
onfig
.R
SSet
.B
aud
$Q
Querying all values of the node "RSSet":
&C
onfig
.R
SSet
$Q
Querying the path of the node "RSSet":
&C
onfig
.R
SSet
$Q.P
Start mode:
&M
ode
$G
Querying the detailed status:
$D