![Metrohm Titrino 719 S Instructions For Use Manual Download Page 74](http://html1.mh-extra.com/html/metrohm/titrino-719-s/titrino-719-s_instructions-for-use-manual_1779645074.webp)
3.2. Remote control commands
719 S Titrino
70
&Info
Object Description
Input
range
Reference
& Root
:
├
I
nfo
Current
data
:
├
.R
eport
Transmission of formatted reports
$G 3.2.2.58.
│
├
.S
elect
Report type
configuration, parameters,
│
smpl data, statistics, silo, calib
│
C-fmla, def, user method,
full
,
│
short, calc, all, ff
ditto
│
├
.C
alibrationData
pH calibration data
$G 3.2.2.59.
│
├
.I
np1
For measuring input 1
│
│
├
.p
Has Asymmetry
pH
0...
7.00
...
±
20.00 ditto
│
│
├
.S
lope
Slope of electrode
0...
1.000
...
±
9.999 ditto
│
│
├
.T
emp Calibration
temperature
-170.0...
25.0
...500.0 ditto
│
│
├
.D
ate
Date of calibration
read only
ditto
│
│
├
.E
lectrodeId
Id of the calibrated electrode
read only
ditto
│
├
.Inp2
For measuring input 2, as for input 1
│
├
.D
iff
For differential input, as for input 1
│
├
.C
hecksums
Checksums
$G 3.2.2.60.
│
├
.A
ctualMethod
Checksum of current method
read only
ditto
│
├
.D
etermData
Determination data
$G 3.2.2.61.
│
├
.W
rite
Read/write for several nods
ON,
OFF
│
├
.E
xV
Volume of Exchange/Dosing unit read
only/read+write
ditto
│
├
.T
itrResults
Titration results
│
├
.R
S
Calculated results
│
│
├
.1
1
st
result
│
│
│
├
.V
alue Value
read
only
3.2.2.62.
│
│
├
up to 9 results
│
├
.E
P
Endpoint
│
│
├
.1
1
st
result
│
│
│
├
.V
Value
read only
ditto
│
│
│
├
.M
eas
Measured value
read only
ditto
│
│
├
up to 2 EP's
│
├
.V
ar
Variables C4X
│
│
├
.C
40
Start measured value
read only/read+write
ditto
│
│
├
.C41
Titration end volume
read only/read+write
│
│
├
.C42
Titration time
read only/read+write
│
│
├
.C43
Volume drift in SET
read only/read+write
│
│
├
.C44
Titration temperature
read only/read+write
│
│
├
.C45
Start volume
read only/read+write
│
│
├
.C46
Asymmetry pH
read only
│
│
├
.C47
Slope of electrode
read only
│
│
├
.D
Time
Time for drift correction read
only/read+write
│
├
.S
tatisticsVal
Statistics values
│
├
.A
ctN
Number of results in chart
read only
3.2.2.63.
│
├
.1
1
st
mean
│
│
├
.M
ean
Mean
read only
ditto
│
│
├
.S
td
Absolute standard deviation
read only
ditto
│
│
├
.R
elStd
Relative standard deviation
read only
ditto
Summary of Contents for Titrino 719 S
Page 2: ......
Page 6: ...1 2 Controls and parts 719 S Titrino 2 1 2 Controls and parts Front view of instrument...
Page 8: ...1 2 Controls and parts 719 S Titrino 4 Rear view of instrument...
Page 58: ......
Page 106: ......