Safety
■■■■■■■■
9
WARNING
Risk of injury by laser radiation
Serious eye injuries by laser radiation.
■
Follow the safety measures and instructions.
■
Instruments must be used by trained personnel only.
■
Instruments of the laser class 3B must be used in protected and
labeled rooms only.
■
The protective glasses provided must be used when working with
open laser beams (e.g. attachment lens).
■
Observe the nominal ocular hazard distance (NOHD).
■
Follow the provisions of the IEC 60825-1 standard "Safety of laser
products" and the regulations for the use of laser systems in your
country.
Laser classification depending on smart tips
Attached smart tip
Laser Class 1
Laser Class 3B
Right Angle Attachment
X
Contact Ball-Probe
X
Calibrate/Verify Attachment (CVA)
X
Vial Holder
X
Short Working Distance Attachment Lens
(SWD)
X
Long Working Distance Attachment Lens
(LWD)
X
Extra Long Working Distance Attachment Lens
(XLWD)
X
Tablet Holder
X
Vial Holder, Tablet Holder and CVA have an interlock mechanism for meas-
urement. This mechanism prevents laser radiation from emerging.
Laser stops immediately if:
■
The lid of the smart tip is opened.
■
Attached smart tip is disconnected from the instrument.