3 Installation
919 IC Autosampler plus
■■■■■■■■
23
3.8.1
Connecting a dosing device
Three dosing devices can be connected to the instrument.
The MagIC Net control software supports only the 800 Dosino as a dosing
device.
■
800 Dosino
WARNING
If a Dosino is connected to the 919 IC Autosampler plus, then the con-
nection cable must be equipped with a T.2400.102 ferrite core. The fer-
rite core diminishes any interference voltages and thus ensures compli-
ance with the strict EMC standards in accordance with the applicable
technical standards, see EU declaration of conformity of the instrument.
Proceed as follows:
1 Mounting the ferrite core
■
Fasten a T.2400.102 ferrite core to the Dosino connection cable
close to the plug.
2 Connecting a dosing device
■
Exit the control software.
■
Connect the connection cable of the dosing device to one of the
sockets marked with
MSB
on the rear of the control device.
■
Start the control software.
USB 1
Contr.
MSB 2
MSB 3
T.2400.102
Figure 13
Connecting a dosing device