![Metrohm Autosampler plus 919 IC Manual Download Page 23](http://html1.mh-extra.com/html/metrohm/autosampler-plus-919-ic/autosampler-plus-919-ic_manual_1779630023.webp)
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
919 IC Autosampler plus
■■■■■■■■
15
Mount the pump tubing as follows:
1 Remove the tubing cartridge
Release the tubing cartridge from the cartridge holder by pressing
the snap-action lever and unhooking it from the mounting bolts
2 Insert the pump tubing
■
Press the contact pressure lever all the way down.
■
Place the pump tubing in the tubing cartridge. The stoppers
must snap into the corresponding holder of the tubing cartridge.
3 Connect the aspiration side
Place a 6.2744.030 tubing olive
on the aspiration side of the
pump tubing.
4 Connect the pressure side
■
of the 6.2744.160 pump tubing connec-
tion (without filter) onto the pump tubing.
■
Select a suitable adapter
(depends on the outer diameter of
the pump tubing) and slide it onto the pump tubing
■
Place the tubing olive
onto the pump tubing.
■
5 Insert the tubing cartridge
■
Hang the tubing cartridge in the mounting bolt and press it
underneath in the cartridge holder until the snap-action lever
snaps in.