![Metrohm 785 DMP Titrino Instructions For Use Manual Download Page 67](http://html1.mh-extra.com/html/metrohm/785-dmp-titrino/785-dmp-titrino_instructions-for-use-manual_1779564067.webp)
2.17. Storing determination results and silo calculations
785 DMP Titrino
64
After several samples have been processed, the silo memory report can have the
following appearance (printout with <PRINT><SILO><ENTER>):
The silo lines can be marked as follows (at very left of report):
+ Silo line has been processed. It cannot be edited anymore.
*
A silo line not yet processed has been deleted.
-
A processed silo line has been deleted and hence removed from the silo
calculations.
/
The last processed silo line. Recalculation will be considered e.g., if the sample data
of this line are changed.
No marking: The silo line is awaiting processing.
For silo lines
≥
100, the first digit will be overwritten by the marking.
2.17.2 Silo calculations
Mean value and standard deviation of the results available in the silo memory can
subsequently be calculated over the entire series.
The following details can be entered in the method under <DEF>:
>silo calculations
C24=
C25=
match id: OFF
Silo calculations
Assignment to C24 and C25
Calculated results (RSX), endpoints (EPX) and
variables (CXX) can be stored as C24.
Identical for C25.
Which sample identifications must match in order to
combine of the results (id1, id1&2, all, OFF)
"OFF" means no matching ids, all samples which have
been processed with the same method are combined,
see examples below.
'si
785 DMP Titrino 0P1/101 785.0010
date 1999-06-27 time 08:54 14
>silo
cycle lines: OFF
save lines: ON
sl method id 1/C21 id 2/C22 id 3/C23 C00 C24 C25
+ 1 11-2 A/12 94-09-12 0.233g 0.142ml/min 98.53%
+ 2 11-2 A/13 94-09-12 0.286g 0.138ml/min 95.75%
/ 3 11-2 A/14 94-09-12 0.197g 0.145ml/min 100.61%
4 11-2 A/15 94-09-12 0.288g NV NV
5 11-2 A/16 94-09-12 0.263g NV NV
processed silo
lines with
saved results
Summary of Contents for 785 DMP Titrino