2.6. Parameters, key <PARAM>
785 DMP Titrino
28
req.smpl size: OFF
cond.
limit smpl size: OFF
cond.
low lim. 0.0
cond.
up lim. 999999
cond.
activate pulse: OFF
cond.
Request of sample size after start of titration (value, unit,
all, OFF)
"all" the value and the unit will be requested.
Limiting value check for sample size (ON, OFF)
With "on" the error message "sample size out." appears
if the entry is outside the set limits. The limiting values
are shown in the display window.
The absolute value of the limit is checked during sam-
ple size input and during the calculation of the results.
If "on" has been set:
Lower limit for sample size (0.0...999 999)
Upper limit for sample size (0.0...999 999)
Pulse output on I/O line 6 (L6, pin 1) of the remote
socket (first, all, cond., OFF)
see page 167.