Technical characteristics of the MTX 3291
6,000- and 60,000-count digital multimeters
55
Technical characteristics of the MTX 3291 (continued)
Resistance
Ohmmeter
In this setting, the user can measure a resistance.
Particular reference conditions:
The (+COM) input must not have been overloaded following the accidental
application of a voltage to the input terminals with the switch set to
Ω
or T°.
If this happens, the return to normal may take about ten minutes.
Protection: 1414 Vpk
Range
Specified
measurement
range
Resolution
Uncertainty
Measurement
current
Open-circuit
voltage
600
Ω
0 to 600.00
Ω
(∗)
0.01
Ω
0.2% L ±20D
≈
1mA
<5V
6k
Ω
0 to 6.0000k
Ω
0.0001k
Ω
0.2% L ±20D
≈
126.6
μ
A
60k
Ω
0 to 60.000k
Ω
0.001k
Ω
≈
12.6
μ
A
600k
Ω
0 to 600.00k
Ω
0.01k
Ω
≈
1.26
μ
A
6M
Ω
0 to 6.0000M
Ω
0.0001M
Ω
1.5% L ±30D
≈
240nA
60M
Ω
0 to 60.000M
Ω
0.001M
Ω
3% L ±30D
≈
29nA
(∗)
REL measurements
Capacity
Capacitance meter
In this setting, the user can measure the capacitance of a capacitor.
Range
Operating
range
Specified
measurement
range
Resolution
Intrinsic
error
Measurement
current
Measurement
time
6nF
0.100 to 6.000nF 0.100 to 6.000nF
0.001nF
2% L ±30D
≈
1.26µA
≈
400ms
60nF
0 to 60.00nF
0 to 60.00nF
0.01nF
1% L ±8D
≈
1.26µA
≈
400ms
600nF 0
to
600.0nF
0 to 600.0nF
0.1nF
1% L ±5D
≈
1.26µA
≈
400ms
6µF
0 to 6.000µF
0 to 6.000µF
0.001µF
1% L ±5D
≈
12.6µA
≈
0.125 s/µF
60µF
0 to 60.00µF
0 to 60.00µF
0.01µF
1% L ±5D
≈
126.6µA
≈
0.125 s/µF
600µF 0
to
600.0µF
0 to 600.0µF
0.1µF
3% L ±5D
≈
1mA
≈
0.125 s/µF
6mF
0 to 6.000mF
0 to 6.000mF
1µF
4% L ±5D
≈
1mA
≈
17 s/mF
60mF
0 to 60.00mF
0 to 60.00mF
10µF
6% L ±5D
≈
1mA
≈
17 s/mF
The use of wires that are very short and shielded is strongly recommended.
Protection: 1414 Vpk