Remote programming
18
100 000 cts digital multimeter
SECondary[?]
(Command/Query)
The command
SEC <group>
selects the group of secondary
measurement.
ATTENTION: All groups of secondary measurements are not valid for
each primary measure.
If the selected group is not valid for the running main measurement,
an error "-221, Settings conflict" is generated.
To the question
SEC?
, the instrument returns the group number of
secondary measurement displayed.
The parameter <group> is an integer numeric value between 0 and
14 that corresponds to the group of secondary measurement
described below:
0:
No secondary measure, graph display.
1:
Frequency, period, mathematical function.
2:
Frequency, dB, mathematical function.
3:
dBm, dBm reference, mathematical function.
4:
Peak positive, negative peak, crest factor.
5:
Power, reference power, mathematical function.
6:
VA, current, mathematical function.
7:
Period, positive duty cycle, mathematical function.
8:
Period, negative duty cycle, mathematical function.
9:
Positive pulse time, positive pulse counter, mathematical
function.
10: Negative pulse time, negative pulse counter, mathematical
function.
11: Mathematical function.
12: Frequency, period.
13: Period, positive duty cycle.
14: Period, negative duty cycle.
TEMPerature:TRANsducer[?]
(Command/Query)
The command
TEMP:TRAN <PT100|PT1000|TCJ|TCK>
selects the
type of transducer used for the temperature measurement.
To the question
TEMP:TRAN?
, the instrument returns the type of
transducer used for the temperature measurement.