Remote programming
100 000 cts digital multimeter
9
*OPC[?]
(Operation complete)
(Command/Query)
To the
*OPC?
question, the instrument returns the value 1 once all pending
commands have been executed.
Query syntax :
*OPC?
Response format :
<1><T>
The
*OPC
command places the message « Operation complete » (bit 0) in
the standard event registry at the end of the running operation.
Command syntax :
*OPC
<value>: Decimal number between 0 and 255 which represents the event
validation registry value.
*RST
(Reset)
(Command)
The
*RST
command resets the instrument in the default factory status.
*SRE[?]
(Service Request
Enable)
(Command/Query)
To the
*SRE?
question, the instrument returns the register value of the
service request activation.
Query syntax :
*SRE?
Response format :
<value><T>
The
*SRE
command sets the service request activation registry.
Command syntax :
*SRE <value>
<value>: Decimal number between 0 and 255 which represents the value of
the service request activation register.
*STB?
(Read Status byte)
(Query)
To the
*STB?
question, the instrument returns the value of the status byte.
Query syntax :
*STB?
Response format :
<value><T>
<value>: Decimal number between 0 and 255 which represents the value of
the status byte.
*TRG
(Trigger)
(Command)
The
*TRG
command sends a trigger demand.
Command syntax
:
*TRG
*TST?
(Self-Test)
(Query)
To the
*TST?
question, the instrument sends a full autoset procedure and
returns the result.
Query syntax :
*TST?
Response format :
<value><T>
<value>: 0 if successful, 1 if failed.
*WAI
(Wait-to-Continue)
(Command)
The
*WAI
command prevents the instrument from performing further
commands as long as the current command has not been terminated.
Command syntax
:
*WAI