Metria COMT-040-001 Instruction Manual Download Page 6

 

Pocket Conductivity Tester 

 

2. 

Align the slot on the new electrode, gently push the electrode into the tester. 

 

 

 

3. 

Twist the electrode collar clockwise until it is tight. Installation is completed. 

 

Troubleshooting 

 

LCD DISPLAY 

CAUSE 

CORRECTIVE ACTION 

 

Electrode dried out 

Soak  the  conductivity  electrode  in  tap  water  for  10 

minutes   

Measured value is out of range 

Check the electrode whether clogged, dirty or broken 

 

Incorrect calibration solution 

Using  the  fresh  conductivity  standard  solutions  for 

calibration 

Setting  value  does  not  match  calibration 

solution 

Reset the calibration value 

Electrode is broken 

Replace the electrode module 

 

Specifications 

 

Model 

COMT-040-001 

COMT-030-001 

COMT-020-001 

Range 

0.1~19.99mS/cm 

10~1999µS/cm 

1.0~199.9µS/cm 

Accuracy 

±1% F.S 

±1% F.S 

±1% F.S 

Resolution 

0.01mS/cm 

1µS/cm 

0.1µS/cm 

Calibration Point 

1 point, 12.88mS/cm 

1 point, 1413µS/cm 

1 point, 146.5µS/cm 

Temperature 

Compensation 

0~50°C, Automatic 

0~50°C, Automatic 

0~50°C, Automatic 

Operating Temperature 

0~60°C, 32~140°F 

0~60°C, 32~140°F 

0~60°C, 32~140°F 

Relative Humidity 

< 80% 

< 80% 

< 80% 

Power Requirements 

3 × 1.5V "G13A" batteries 

3 × 1.5V "G13A" batteries 

3 × 1.5V "G13A" batteries 

Dimensions 

185 (L) × 40 (Dia.) mm 

185 (L) × 40 (Dia.) mm 

185 (L) × 40 (Dia.) mm 

Weight 

100g 

100g 

100g 

 