Metria COMT-040-001 Instruction Manual Download Page 4

Pocket Conductivity Tester 

 

Conductivity Calibration 

Pocket conductivity tester allows 1 point calibration. To ensure accurate measurement, we recommend that calibrating the meter regularly. 
The tester acceptable calibration ranges from 10 to 18mS/cm (COMT-040-001), 1000 to 1800 µS/cm (COMT-030-001) and 100 to 180 µS/cm 
(COMT-020-001). Make sure that you use fresh conductivity standard solution during the calibration. DO NOT reuse calibration solution as it 
may be contaminated and affect the calibration and accuracy of measurement. 
 

 

1. 

Rinse  the  conductivity  electrode  with  distilled  water  ,  then  rinse  with  a  small 

amount of standard solution. 
2. 

Place  the  electrode  into  the  standard  solution  (e.g.,  12.88mS/cm),  stir  the  tester 

gently. Make sure there are no air bubble trapped in the slot of the sensor. 

 

3. 

Press and hold the 

Cal

 key, the tester enters the calibration mode. 

 

4. 

Press  the 

  or 

  key  to  set  the  calibration  value.  Ensure  the  displayed  value 

matches your   

calibration standard. 

 
 

 

 

Press 

 or 

 key once, the setting value will gradually increase or decrease. Press 

and hold    the 

 or 

 key, the setting value will quickly increase or decrease. 

 

 

5. 

Press the 

Meas

 key, the calibration value will automatically flash 3 times. The tester 

returns to the measurement mode. Calibration is completed. 

 

Exit the calibration 

During the calibration process, if you want to exit the calibration mode, DO NOT press the 

Meas

 key in the step 5. Press the 

 or 

 key 

until the display shows “

---

“, then press 

the

 Meas

 key to exit. 

 

 
 
 
 