MI 3108 EurotestPV
Measurements - Power & energy
88
Connection for current measurement
Figure 7.9: Leakage and load current measurements
Current measurement procedure
Select
CURRENT
function using the function selector keys.
Select input channel (optional).
Connect
current clamp(s) to the instrument.
Connect
the clamp(s) to the item to be tested (see figure 7.9).
Press the
TEST
key to start the continuous measurement.
Press the
TEST
key again to stop the measurement.
Store
the result by pressing the
MEM
key (optional).
Figure 7.10: Examples of current measurement result
Displayed results for Current measurement:
I, I1, I2 …..…..Current
Note
:
Channel
C2 is intended for measuring with clamps A 1391 only.