MI 3108 EurotestPV
Appendix D
– Commanders
122
Figure D.2: Back side
Legend:
1 TEST
TEST
Starts measurements.
Acts also as the PE touching electrode.
2 LED
Left status RGB LED
3 LED
Right status RGB LED
4 LEDs
Lamp LEDs (Tip commander)
5 Function selector Selects test function.
6 MEM
Store / recall / clear tests in memory of instrument.
7 BL
Switches On/Off backlight on instrument
8 Lamp key
Switches On/ Off lamp (Tip commander)
9 Battery cells
Size AAA, alkaline / rechargeable NiMH
10 Battery cover
Battery compartment cover
11 Cap
Removable CAT IV cap (Tip commander)
D.4 Operation of commander
Both LED yellow
Warning!
Dangerous voltage on the commander’s
PE terminal!
Right LED red
Fail indication
Right LED green
Pass indication
Left LED blinks blue
Commander is monitoring the input voltage
Left LED orange
Voltage between any test terminals is higher than
50 V
Both LEDs blink red
Low battery
Both LEDs red and switch off Battery voltage too low for operation of commander