A 1179, A 1257, A 1395 Flex current clamps 
 

Procedure 

Notes 

 

Completely 

charge

 the battery. 

Time depends on the battery capacity 
(10 – 24 hours)  

 

Completely 

discharge

 the battery. 

Can be performed with normal work with 
the A1179/A1257/A1395. 

 

Repeat

 the charge / discharge cycle 

for at least 

two times

Four cycles are recommended. 

 
Complete discharge / charge cycle is performed automatically for each cell using 
external intelligent battery charger. 
 

Notes: 

 

The charger in the A1179/A1257/A1395 is a pack cell charger. This means that 
the battery cells are connected in series during the charging. The battery cells 
have to be equivalent (same charge condition, same type and age). 

 

One different battery cell can cause an improper charging and incorrect 
discharging during normal usage of the entire battery pack (it results in heating of 
the battery pack, significantly decreased operation time, reversed polarity of 
defective cell,…). 

 

If no improvement is achieved after several charge / discharge cycles, then each 
battery cell should be checked (by comparing battery voltages, testing them in a 
cell charger, etc). It is very likely that only some of the battery cells are 
deteriorated. 

 

The effects described above should not be mixed with normal decrease of 
battery capacity over time. Battery also loses some capacity when it is 
repeatedly charged / discharged. Actual decreasing of capacity, versus number 
of charging cycles, depends on battery type. It is provided in the technical 
specification from battery manufacturer. 

 

6.2 User interface 

 

Power LED

Range LEDs 

{

 KEY 

 

 

Picture 6.2: Front user interface view for A1179 

Summary of Contents for A 1179

Page 1: ...amps A 1179 A 1257 A 1395 User Manual Version 1 6 Code No 20 751 167 KOMETEC Karl Oelkers e K Mess und Pr fger te Shop Mozartstr 10 D 88097 Eriskirch T 07541 955 1313 F 07541 955 1131 info kometec de...

Page 2: ...www metrel si Mark on your equipment certifies that this equipment meets the requirements of the EC European Community regulations concerning safety and electromagnetic compatibility 2007 2008 2012 Me...

Page 3: ...A1257 A1395 Flexible Current Clamps 7 6 1 Battery and charging 7 6 1 1 New battery cells or cells unused for a longer period 7 6 2 User interface 8 6 2 1 A1179 A1257 A1395 states regarding power supp...

Page 4: ...ut is not currently present the user can reset the blinking Range LED by pressing the key If the overrange is present also when the key is pressed the range will be changed to the less sensitive range...

Page 5: ...insulation Also look for loose or weakened components Do not use a clamp that is cracked damaged or has a defective cable Never use the clamp on a circuit with voltages higher than 1000 V CAT III or...

Page 6: ...with the relevant European standards 5 Output connecting The A1179 A1257 A1395 Flexible Current Clamps can be used with any Metrel Power Analyzer Oscilloscope or Multimeter that has the following fea...

Page 7: ...ed to the clamps and power off the A1179 A1257 Insert cells correctly otherwise the A1179 A1257 A1395 will not operate and the battery could be discharged Remove all battery cells from the battery com...

Page 8: ...cause an improper charging and incorrect discharging during normal usage of the entire battery pack it results in heating of the battery pack significantly decreased operation time reversed polarity...

Page 9: ...A1395 module states regarding power supply and charging and actions to change between them are defined State How is it indicating What is happening POWER OFF All LED diodes OFF External power supply i...

Page 10: ...ge is too small for measured current amplitude peak value If the overrange current was detected but currently it is not present any more the user can reset blinking Range LED by pressing the key see T...

Page 11: ...verrange OVERRANGE DETECTION 2000 3000 A OVERRANGE DETECTION 2000 3000 A Press the key RANGE 2000 3000 A Table 6 5 Change state table when single overrange is detected State before Action to change st...

Page 12: ...pling See Picture 7 1 5 It is very important that the conductor is as the center and perpendicular to the measuring head if it is no possible additional error of 2 of full scale can occur 6 Do not mea...

Page 13: ...it Do not use liquids based on petrol or hydrocarbons Do not spill cleaning liquid over the A1179 A1257 A1395 9 Service For repairs under or out of warranty time please contact your Metrel distributor...

Page 14: ...l output voltage 1 V at full scale Minimum load impedance in output 2 10 k internal output is 150 Phase error 1 50Hz 10 1 5kHz Crest factor Up to 3 0 Overrange indication Corresponding range LED indic...

Page 15: ...nce conditions 4 Expression of error measurement in refence conditions error Iresidual I measured x 100 linearity Accuracy 5 Full charged NIMiH 2200 mAh at 25 C 6 Distance between cable and clamp coup...

Reviews: