A 1179, A 1257, A 1395 Flex current clamps 
 

Range LEDs: 

one of these LEDs indicates corresponding measuring range of the 

output signal. If the LED is blinking (1 s ON – 1s OFF) 
A1179/A1257/A1395 detect overrange event at the currently selected 
range. With pressing the KEY one can reset overrange detection. If all 
Range LEDs are OFF it indicates power off the A1179/A1257/A1395. 

Power LED: 

OFF – normal operation 
Blinking (1 s ON – 1s OFF) – low battery indication, the battery is too 
weak for correct measurement results. 
ON – The battery cells are charged. 

KEY: 

Turns the A1179/A1257/A1395 ON, change the measuring range, reset 
overrange detection, if held in for more then 3 seconds it turns the 
A1179/A1257/A1395 module OFF.

 

 
See detailed description of A1179/A1257/A1395 operating states in chapter 6.2.1 and 
6.2.2. 
 

6.2.1 A1179/A1257/A1395 states regarding power supply 

 

The use of the A1179/A1257/A1395 module is very easy and intuitive. In 

following tables (Table 6.1 and Table 6.2) the A1179/A1257/A1395 module states 
regarding power supply and charging and actions to change between them are defined.   
 
State 

How is it indicating? 

What is happening? 

POWER OFF 

All LED diodes OFF, External 
power supply is not connected. 

A1179/A1257/A1395 is not 
working. Battery is not charged. 

NORMAL 

One of the Range LED is ON, Low 
bat is OFF . 

A1179/A1257/A1395 is working 
from the battery. Battery is not 
charged. 

EXTERNAL 
NORMAL 

Low bat LED is ON, External 
power supply is connected and 
batteries 

are not present.

. One of 

the Range LED is ON. 

A1179/A1257/A1395 is working 
from the external power supply. 
Battery is not charged.   

LOW 
BATTERY 

Low bat LED is blinking. 

A1179/A1257/A1395 is working. 
Battery is low, Measuring results 
may not be in specifications. 
Battery is not charged. 

 
NORMAL 

One of the Range LED is ON, Low 
bat is ON External power supply is 
connected and rechargable 
batteries are present.  

Battery is charged. 
A1179/A1257/A1395 is powered 
from external power supply (not 
from batteries). 

CHARGE 

All Range LEDs are OFF. Low bat 
is ON External power supply is 
connected and rechargable 
batteries are present. 

Battery is charged. No 
measuring is possible. 

Table 6.1 Power supply state indications 

 
 
 

Summary of Contents for A 1179

Page 1: ...amps A 1179 A 1257 A 1395 User Manual Version 1 6 Code No 20 751 167 KOMETEC Karl Oelkers e K Mess und Pr fger te Shop Mozartstr 10 D 88097 Eriskirch T 07541 955 1313 F 07541 955 1131 info kometec de...

Page 2: ...www metrel si Mark on your equipment certifies that this equipment meets the requirements of the EC European Community regulations concerning safety and electromagnetic compatibility 2007 2008 2012 Me...

Page 3: ...A1257 A1395 Flexible Current Clamps 7 6 1 Battery and charging 7 6 1 1 New battery cells or cells unused for a longer period 7 6 2 User interface 8 6 2 1 A1179 A1257 A1395 states regarding power supp...

Page 4: ...ut is not currently present the user can reset the blinking Range LED by pressing the key If the overrange is present also when the key is pressed the range will be changed to the less sensitive range...

Page 5: ...insulation Also look for loose or weakened components Do not use a clamp that is cracked damaged or has a defective cable Never use the clamp on a circuit with voltages higher than 1000 V CAT III or...

Page 6: ...with the relevant European standards 5 Output connecting The A1179 A1257 A1395 Flexible Current Clamps can be used with any Metrel Power Analyzer Oscilloscope or Multimeter that has the following fea...

Page 7: ...ed to the clamps and power off the A1179 A1257 Insert cells correctly otherwise the A1179 A1257 A1395 will not operate and the battery could be discharged Remove all battery cells from the battery com...

Page 8: ...cause an improper charging and incorrect discharging during normal usage of the entire battery pack it results in heating of the battery pack significantly decreased operation time reversed polarity...

Page 9: ...A1395 module states regarding power supply and charging and actions to change between them are defined State How is it indicating What is happening POWER OFF All LED diodes OFF External power supply i...

Page 10: ...ge is too small for measured current amplitude peak value If the overrange current was detected but currently it is not present any more the user can reset blinking Range LED by pressing the key see T...

Page 11: ...verrange OVERRANGE DETECTION 2000 3000 A OVERRANGE DETECTION 2000 3000 A Press the key RANGE 2000 3000 A Table 6 5 Change state table when single overrange is detected State before Action to change st...

Page 12: ...pling See Picture 7 1 5 It is very important that the conductor is as the center and perpendicular to the measuring head if it is no possible additional error of 2 of full scale can occur 6 Do not mea...

Page 13: ...it Do not use liquids based on petrol or hydrocarbons Do not spill cleaning liquid over the A1179 A1257 A1395 9 Service For repairs under or out of warranty time please contact your Metrel distributor...

Page 14: ...l output voltage 1 V at full scale Minimum load impedance in output 2 10 k internal output is 150 Phase error 1 50Hz 10 1 5kHz Crest factor Up to 3 0 Overrange indication Corresponding range LED indic...

Page 15: ...nce conditions 4 Expression of error measurement in refence conditions error Iresidual I measured x 100 linearity Accuracy 5 Full charged NIMiH 2200 mAh at 25 C 6 Distance between cable and clamp coup...

Reviews: