5
CLEANING AND MAINTENANCE
5.2.2
Thermopile Sensor Cleaning
The lens of this sensor must be free of all dirt and lint to be accurate.
1. WASH–using a Kimwipe tissue moistened with cleaning solu-
tion or isopropyl alcohol to clean the thermopile sensor.
2. RINSE–using a new Kimwipe moistened with steam distilled
water to rinse the cleaning solution from the sensor.
3. DRY–use a dry Kimwipe to help remove any moisture remain-
ing from the cleaning process.
5.2.3
Chamber Cleaning Instructions
Wrap a new Kimwipe strip around the end of the swab (included in
the cleaning kit) and moisten it with cleaning solution or isopropyl
alcohol.
1. WASH–clean the surrounding chamber area with the moist
Kimwipe. The chamber area, especially where the cup seals,
must be free of all contamination.
2. RINSE–repeat the steps above using a new Kimwipe strip moist-
ened with steam distilled water.
3. DRY–repeat steps above again, this time using a dry Kimwipe
strip to remove any moisture remaining from the cleaning pro-
cess.
Note: Do NOT reuse Kimwipes.
23