PN T900-014, Rev B
Actuator Gas
Specification
Description
Inlet Fitting
1/8-inch VICI compression fitting
Inlet Pressure Range
4.8–10.2 BAR (70–150 PSIG)
Gas Type
Inerts (Nitrogen, Argon, Helium), 7 nines or better or CDA,
dewpoint < -65 °C
Gas Type
(for ta7000F-O
2
H
2
only)
Inerts (Nitrogen, Argon, Helium), 7 nines or better
Calibration Gas
Specification
Description
Inlet Fitting
1/16-inch VICI compression fitting
Cylinder Concentration
5–10 PPM H
2
, CO, CH
4
, CO
2
, and Ethane (C
2
H
6
) balance Nitrogen
Cylinder Volume
100 cubic feet minimum (2800 L)
Supply Pressure
5.3–6.1 BAR (80–90 PSIG)
Consumption
Approximately 1.0 L per calibration (4–8 L/month), automatic
shut-off.
FID Hydrogen
Specification
Description
Inlet Fitting
1/16-inch VICI compression fitting
Inlet Pressure Range
3.4–4.1 BAR (50–60 PSIG)
Inlet Pressure Stability
+2 %, UHP regulator required
Flow Rate
35 cc/minute
Purity
Hydrocarbons, CO, CO, Moisture <1 PPM
FID Air
Specification
Description
Inlet Fitting
1/16-inch VICI compression fitting
Inlet pressure range
3.4–4.1 BAR (50–60 PSIG)
Inlet pressure stability
+2 %, regulator required
Flow rate
250 cc/minute
Purity
Total hydrocarbons <1 PPM, dewpoint < -65° C
Specifications
|