5 - 11
LWIR
Thermal Imaging Camera
Installation Guide
Input/Output 3
For
mA In
, use the drop-down boxes and text input fields to the set the
required values for this input.
For example, you could set a
Background Temperature
input for
ROI 1
where
4 mA
corresponds to 0 °C and
20 mA
corresponds to 50 °C.
For Input/Output 3, the
Relay
settings are identical in operation to those of
Output 2.
The options available for the Command input (
Cmd In
) for Input/Output 3 each
relate to the LWIR
Camera
and are as follows:
•
NUC
- the command input will trigger a Non-Uniformity Correction in the
Camera;
•
Auto NUC Disable
- the command input will switch off the automatic
Non-Uniformity Correction in the Camera;
•
Reset Time Function
- The command input will reset any time functions
that are active e.g. Peak Picker, Averager, Track and Hold etc.
Summary of Contents for LAND LWIR