BT-610-9800 Rev C
11. Specifications
Performance
Particle Size Range
0.3µm - 10µm, 6 channels
Calibrated Sizes
0.3 µm, 0.5µm, 1.0µm, 2.0µm 5.0µm and 10µm
User Size Settings
0.1µm steps from 0.3µm - 2.0µm
0.5µm steps from 2.0µm - 10µm
Concentration Range
0
– 3,000,000 particles per cubic foot (105,900 particle/L)
Accuracy
± 10% to calibration aerosol
Sensitivity
0.3 µm
Flow Rate
0.1 cfm (2.83 lpm)
Sample Time
Adjustable: 1 to 9999 seconds
Hold Time
Adjustable: 0 to 9999 seconds
Electrical
Light Source
Laser Diode, 90mW, 780 nm
Power
12V Li-Ion self-contained battery pack
– Provides for 8 hours
continuous use. Full recharge approximately 3 hours.
AC Adapter/Charger
AC to DC module, 100
– 240 VAC to 16.8 VDC @ 1.8 A
Printer
21 character per line thermal printer
Communications
USB, RS-232 or RS-485
Standards
Meets ISO 21501-4
Interface
Display
20 character x 4 line LCD
Keyboard
8 key membrane type
Physical
Height
7
” (17.7 cm)
Width
6.6” (16.8 cm)
Depth
6.5
” (16.5 cm)
Weight
5.00 lbs
– 80 ounces – (2.27 kg)
Environmental
Operating Temperature
0
º
C to +50º C
Storage Temperature
-20º C to +60º C
Accessories
Supplied
Operation Manual
(P/N BT-610-9800)
USB Cable
(P/N 500784)
Comet Software
(P/N 80248)
Particle View Software
Particle View
Battery Charger
(P/N 80643)
Iso-kinetic Sample Probe
(P/N G3110)
Zero Particulate Filter
(P/N G3111)
Printer Paper (2 Rolls)
(P/N 750514)
Optional
RH & Temperature Probe
(P/N G3120)
Flow Meter
(P/N 9801)
Serial Cable