2. Inhalation
Hold the nebulizer in your hand stably and start inhalation.
Attention
Do not cover the vent when you inhale. Otherwise, that will decrease
the nebulizationrate.
Attention
• If the device detects no medication in medication chamber, it will
shut off automatically.
• If the device does not shut off automatically when medication
depleted , press the “START/STOP” button to turn the power off
immediately to avoid the mesh breaking. Please go to page 18 for
troubleshooting.
• During the treatment, you may adjust the nebulizer to any an
-
gle. However, make sure the medication stays in contact with the
mesh; otherwise, the nebulizer will shut off automatically after
approximately 15seconds.
• When the medication is about to be depleted, it is recommended
that you tilt the nebulizer (the buttons side) slightly toward you.
This allows the remaining medication to contact the mesh to ne
-
bulization.
• Do not shake the nebulizer strongly in the usage. Otherwise, the
nebulizer may shut off automatically.
• Provide close supervision when the nebulizer is used by children.